Produkty
Specifications |
Q64RD |
Q64RD-G |
Q64TD |
Q64TDV-GH |
Q68RD3-G |
Q68TD-G-H01/H02 |
---|---|---|---|---|---|---|
Input channels |
4 |
4 |
4 |
4 |
8 |
8 |
Connectable temperature sensors type |
Pt100 (conforms to JIS C 1604-1989 and DIN IEC 751), JPt100 (conforms to JIS C 1604-1981) |
Pt100 (conforms to JIS C 1604-1997 and DIN IEC 751-1983), JPt100 (conforms to JIS C 1604-1981), Ni100Ω (conforms to DIN 43760-1987) |
K, E, J, T, B, R, S, N (conforms to JIS C1602-1995, IEC 584-1 and 584-2) |
K, E, J, T, B, R, S, N (conforms to JIS C1602-1995, IEC 584-1 and 584-2) |
Pt100 (conforms to JIS C 1604-1997 and DIN IEC 751), JPt100 (conforms to JIS C 1604-1981), Ni100Ω (conforms to DIN 43760-1987) |
K, E, J, T, B, R, S, N (conforms to JIS C1602-1995, IEC 584-1 and 584-2) |
Temperature measuring range |
Pt100: -200–850 °C, JPt 100: -180–600 °C |
Pt100: -200–850 °C, JPt100: -180–600 °C, Ni100 Ω: -60–180 °C |
Depends on the thermocouple used |
Depends on the thermocouple used |
Pt100: -200–850 °C, JPt100: -180–600 °C, Ni100Ω: -60–180 °C |
Depends on the thermocouple used |
Temperature scaling value |
16-bit, signed binary: -2000–8500 32-bit, signed binary: -200 000–850 000 |
16-bit, signed binary: -2000–8500 32-bit, signed binary: -200 000–850 000 |
16-bit, signed binary: -2700–18 200 32-bit, signed binary: — |
16-bit, signed binary: -25 000–25 000 32-bit, signed binary: — |
16-bit, signed binary: -2000–8500 |
16-bit, signed binary: -2700–18 200 |
Max. resolution °C |
0.025 °C |
0.025 °C |
B, R, S, N: 0.3 °C; K, E, J, T: 0.1 °C |
B: 0.7 °C; R, S: 0.8 °C; K, T: 0.3 °C; E: 0.2 °C; J: 0.1 °C; N: 0.4 °C; Voltage: 4 µV |
0.1 °C |
B, R, S, N: 0.3 °C; K, E, J, T: 0.1 °C |
Cold junction temp. compensation accuracy |
— |
— |
±1.0 °C |
±1.0 °C |
— |
provided |
Overall accuracy |
±0.08 % (accuracy relative to full-scale value) at ambient temperature 25±5 °C |
±0.04 % (accuracy relative to full-scale value) at ambient temperature 25±5 °C |
Depends on the thermocouple used |
Depends on the thermocouple used |
Depends on the thermocouple used |
Depends on the thermocouple used |
Max. conversion time |
40 ms/channel |
40 ms/channel |
20 ms/channel |
20 ms/channel |
320 ms/8 channels |
320 ms/8 channels (H01), 640 ms/8 channels (H02) |
Analog inputs |
4 channels/module |
4 channels/module |
4 channels/module + Pt100 connection |
4 channels/module + Pt100 connection |
8 channels |
8 channels/module |
Temp. measurement output current mA |
1 |
1 |
— |
— |
1 |
— |
Insulation method |
Transformer insulation 1 |
Photocoupler insulation 2 Transformer insulation 3 |
Transformer insulation 4 |
Transformer insulation 5 |
Transformer insulation 5 |
Transformer insulation 5 |
Disconnection detection |
For each channel independent | |||||
Occupied I/O points |
16 |
16 |
16 |
16 |
16 |
16 |
Connection terminal |
All modules are fitted with a removable terminal block with 18 screw terminals. |
A6CON 40pin connector | ||||
Applicable wire size mm 2 |
0.3–0.75 |
0.3–0.75 |
0.3–0.75 |
0.3–0.75 |
≤0.3 |
≤0.3 |
Internal power consumption (5 V DC) mA |
600 |
620 |
500 |
500 |
0.54 A |
0.49 A (H01) 0.65 A (H02) |
Weight kg |
0.17 |
0.20 |
0.25 |
0.25 |
0.20 |
0.17 |
Dimensions (WxHxD) mm |
27.4x98x90 |
27.4x98x112 |
27.4x98x90 |
27.4x98x90 |
27.4x102x130 |
27.4x98x90 (H01) 27.4x102x130 (H02) |
Order information Art. no. |
137592 |
154749 |
137591 |
143544 |
216482 |
216481/221582 |
1 between power supply and temperature inputs 2 between each channel and PLC power 3 between measuring input channels 4 between thermocouple inputs as well as thermocouple and earth 5 between each channel and between the channels and PLC power
Specifications |
Q64RD |
Q64RD-G |
Q64TD |
Q64TDV-GH |
Q68RD3-G |
Q68TD-G-H01/H02 |
---|---|---|---|---|---|---|
Input channels |
4 |
4 |
4 |
4 |
8 |
8 |
Connectable temperature sensors type |
Pt100 (conforms to JIS C 1604-1989 and DIN IEC 751), JPt100 (conforms to JIS C 1604-1981) |
Pt100 (conforms to JIS C 1604-1997 and DIN IEC 751-1983), JPt100 (conforms to JIS C 1604-1981), Ni100Ω (conforms to DIN 43760-1987) |
K, E, J, T, B, R, S, N (conforms to JIS C1602-1995, IEC 584-1 and 584-2) |
K, E, J, T, B, R, S, N (conforms to JIS C1602-1995, IEC 584-1 and 584-2) |
Pt100 (conforms to JIS C 1604-1997 and DIN IEC 751), JPt100 (conforms to JIS C 1604-1981), Ni100Ω (conforms to DIN 43760-1987) |
K, E, J, T, B, R, S, N (conforms to JIS C1602-1995, IEC 584-1 and 584-2) |
Temperature measuring range |
Pt100: -200–850 °C, JPt 100: -180–600 °C |
Pt100: -200–850 °C, JPt100: -180–600 °C, Ni100 Ω: -60–180 °C |
Depends on the thermocouple used |
Depends on the thermocouple used |
Pt100: -200–850 °C, JPt100: -180–600 °C, Ni100Ω: -60–180 °C |
Depends on the thermocouple used |
Temperature scaling value |
16-bit, signed binary: -2000–8500 32-bit, signed binary: -200 000–850 000 |
16-bit, signed binary: -2000–8500 32-bit, signed binary: -200 000–850 000 |
16-bit, signed binary: -2700–18 200 32-bit, signed binary: — |
16-bit, signed binary: -25 000–25 000 32-bit, signed binary: — |
16-bit, signed binary: -2000–8500 |
16-bit, signed binary: -2700–18 200 |
Max. resolution °C |
0.025 °C |
0.025 °C |
B, R, S, N: 0.3 °C; K, E, J, T: 0.1 °C |
B: 0.7 °C; R, S: 0.8 °C; K, T: 0.3 °C; E: 0.2 °C; J: 0.1 °C; N: 0.4 °C; Voltage: 4 µV |
0.1 °C |
B, R, S, N: 0.3 °C; K, E, J, T: 0.1 °C |
Cold junction temp. compensation accuracy |
— |
— |
±1.0 °C |
±1.0 °C |
— |
provided |
Overall accuracy |
±0.08 % (accuracy relative to full-scale value) at ambient temperature 25±5 °C |
±0.04 % (accuracy relative to full-scale value) at ambient temperature 25±5 °C |
Depends on the thermocouple used |
Depends on the thermocouple used |
Depends on the thermocouple used |
Depends on the thermocouple used |
Max. conversion time |
40 ms/channel |
40 ms/channel |
20 ms/channel |
20 ms/channel |
320 ms/8 channels |
320 ms/8 channels (H01), 640 ms/8 channels (H02) |
Analog inputs |
4 channels/module |
4 channels/module |
4 channels/module + Pt100 connection |
4 channels/module + Pt100 connection |
8 channels |
8 channels/module |
Temp. measurement output current mA |
1 |
1 |
— |
— |
1 |
— |
Insulation method |
Transformer insulation 1 |
Photocoupler insulation 2 Transformer insulation 3 |
Transformer insulation 4 |
Transformer insulation 5 |
Transformer insulation 5 |
Transformer insulation 5 |
Disconnection detection |
For each channel independent |
For each channel independent |
For each channel independent |
For each channel independent |
For each channel independent |
For each channel independent |
Occupied I/O points |
16 |
16 |
16 |
16 |
16 |
16 |
Connection terminal |
All modules are fitted with a removable terminal block with 18 screw terminals. |
All modules are fitted with a removable terminal block with 18 screw terminals. |
All modules are fitted with a removable terminal block with 18 screw terminals. |
All modules are fitted with a removable terminal block with 18 screw terminals. |
A6CON 40pin connector |
A6CON 40pin connector |
Applicable wire size mm 2 |
0.3–0.75 |
0.3–0.75 |
0.3–0.75 |
0.3–0.75 |
≤0.3 |
≤0.3 |
Internal power consumption (5 V DC) mA |
600 |
620 |
500 |
500 |
0.54 A |
0.49 A (H01) 0.65 A (H02) |
Weight kg |
0.17 |
0.20 |
0.25 |
0.25 |
0.20 |
0.17 |
Dimensions (WxHxD) mm |
27.4x98x90 |
27.4x98x112 |
27.4x98x90 |
27.4x98x90 |
27.4x102x130 |
27.4x98x90 (H01) 27.4x102x130 (H02) |
Order information Art. no. |
137592 |
154749 |
137591 |
143544 |
216482 |
216481/221582 |
1 between power supply and temperature inputs 2 between each channel and PLC power 3 between measuring input channels 4 between thermocouple inputs as well as thermocouple and earth 5 between each channel and between the channels and PLC power